Source:Power#

SCPI Commands

[SOURce<HW>]:POWer:AATTenuation
[SOURce<HW>]:POWer:IQPep
[SOURce<HW>]:POWer:LBEHaviour
[SOURce<HW>]:POWer:LMODe
[SOURce<HW>]:POWer:MANual
[SOURce<HW>]:POWer:MODE
[SOURce<HW>]:POWer:PEP
[SOURce<HW>]:POWer:POWer
[SOURce<HW>]:POWer:STARt
[SOURce<HW>]:POWer:STOP
[SOURce]:POWer:WIGNore

Cloning the Group

// Create a clone of the original group, that exists independently
var power2 = driver.Source.Power.Clone();

Direct SCPI Members

double Aattenuation#
// SCPI: [SOURce<HW>]:POWer:AATTenuation
double value = driver.Source.Power.Aattenuation;

Queries the currently active level attenuation.

activeAtt: float


double IqPep#
// SCPI: [SOURce<HW>]:POWer:IQPep
double value = driver.Source.Power.IqPep;

No help available.


PowLevBehaviourEnum Lbehaviour#
// SCPI: [SOURce<HW>]:POWer:LBEHaviour
PowLevBehaviourEnum value = driver.Source.Power.Lbehaviour;
driver.Source.Power.Lbehaviour = PowLevBehaviourEnum.AUTO;

Selects the level behavior at the RF output over time.

behaviour: AUTO | UNINterrupted | MONotone | CVSWr | CPHase

  • UNINterrupted | MONotone: Do not use the uninterrupted level settings and strictly monotone modes in combination with the high-quality optimization mode (see [:SOURcehw]:BB:IMPairment:OPTimization:MODE)

  • CVSWr: Constant VSWR

  • CPHase: Continuous phase

  • CONStant: Constant phase


PowLevModeEnum Lmode#
// SCPI: [SOURce<HW>]:POWer:LMODe
PowLevModeEnum value = driver.Source.Power.Lmode;
driver.Source.Power.Lmode = PowLevModeEnum.LOWDistortion;

Sets the RF level mode.

levMode: NORMal | LOWNoise | LOWDistortion

  • NORMal: Supplies the RF signal with the standard power level of the instrument.

  • LOWNoise: Supplies a very low noise sinewave signal.

  • LOWDistortion: Supplies a very pure sinewave signal.


double Manual#
// SCPI: [SOURce<HW>]:POWer:MANual
double value = driver.Source.Power.Manual;
driver.Source.Power.Manual = 1.0;

Sets the level for the subsequent sweep step if method RsSmw.Source.Sweep.Power.Mode. Use a separate command for each sweep step.

manual: You can select any level within the setting range, where: STARt is set with method RsSmw.Source.Power.Start() STOP is set with method RsSmw.Source.Power.Stop() OFFSet is set with method RsSmw.Source.Power.Level.Immediate.Offset()


LfFreqModeEnum Mode#
// SCPI: [SOURce<HW>]:POWer:MODE
LfFreqModeEnum value = driver.Source.Power.Mode;
driver.Source.Power.Mode = LfFreqModeEnum.CW;

Selects the operating mode of the instrument to set the output level.

mode: CW | FIXed | SWEep

  • CW | FIXed: Operates at a constant level.CW and FIXed are synonyms.To set the output level value, use the command [:SOURcehw]:POWer[:LEVel][:IMMediate][:AMPLitude].

  • SWEep: Sets sweep mode.Set the range and current level with the commands:[:SOURcehw]:POWer:STARt and [:SOURcehw]:POWer:STOP,[:SOURcehw]:POWer:MANual.


double Pep#
// SCPI: [SOURce<HW>]:POWer:PEP
double value = driver.Source.Power.Pep;

Queries the PEP “Peak Envelope Power) of digital modulation or digital standards at the RF output. This value corresponds to the level specification, displayed in the status bar (header) .

pep: float


double Power#
// SCPI: [SOURce<HW>]:POWer:POWer
double value = driver.Source.Power.Power;
driver.Source.Power.Power = 1.0;

Sets the level at the RF output connector. This value does not consider a specified offset. The command method RsSmw.Source.Power.Level.Immediate.Amplitude sets the level of the Level display, that means the level containing offset. See “Displayed RF frequency and level values with downstream instruments”.

power: Level at the RF output, without level offset


double Start#
// SCPI: [SOURce<HW>]:POWer:STARt
double value = driver.Source.Power.Start;
driver.Source.Power.Start = 1.0;

Sets the RF start/stop level in sweep mode.


double Stop#
// SCPI: [SOURce<HW>]:POWer:STOP
double value = driver.Source.Power.Stop;
driver.Source.Power.Stop = 1.0;

Sets the RF start/stop level in sweep mode.

stop: Sets the setting range calculated as follows: (Level_min + OFFSet) to (Level_max + OFFSet) Where the values are set with the commands: method RsSmw.Source.Power.Level.Immediate.Offset() method RsSmw.Source.Power.Start() method RsSmw.Source.Power.Stop()


bool Wignore#
// SCPI: [SOURce]:POWer:WIGNore
bool value = driver.Source.Power.Wignore;
driver.Source.Power.Wignore = false;

Ignores level range warnings.

state: 1 | ON | 0| OFF


Subgroups