Source:Power:Alc#
SCPI Commands
- [SOURce<HW>]:POWer:ALC:DAMPlifier
- [SOURce<HW>]:POWer:ALC:DSENsitivity
- [SOURce<HW>]:POWer:ALC:MODE
- [SOURce<HW>]:POWer:ALC:OMODe
- [SOURce<HW>]:POWer:ALC:SEARch
- [SOURce<HW>]:POWer:ALC:SLEVel
- [SOURce<HW>]:POWer:ALC:[STATe]
Cloning the Group
// Create a clone of the original group, that exists independently
var alc2 = driver.Source.Power.Alc.Clone();
Direct SCPI Members
- PowAlcDriverAmpEnum Damplifier#
// SCPI: [SOURce<HW>]:POWer:ALC:DAMPlifier PowAlcDriverAmpEnum value = driver.Source.Power.Alc.Damplifier; driver.Source.Power.Alc.Damplifier = PowAlcDriverAmpEnum.AUTO;Selects the driver amplifier switching state.
amplifier: OFF | ON | AUTO | FIX | ONMG
AUTO: Switches the attenuator automatically.
ON | OFF: Switches on or off the driver amplifier.
FIXed: Fixes the last setting.
ONMG: Supplies maximum level at the output.
- PowAlcDetSensitivityEnum Dsensitivity#
// SCPI: [SOURce<HW>]:POWer:ALC:DSENsitivity PowAlcDetSensitivityEnum value = driver.Source.Power.Alc.Dsensitivity; driver.Source.Power.Alc.Dsensitivity = PowAlcDetSensitivityEnum.AUTO;Sets the sensitivity of the ALC detector.
sensitivity: AUTO | FIXed | LOW | MEDium | HIGH
AUTO: Selects the optimum sensitivity automatically.
LOW | MEDium | HIGH: Sets either low, medium or high sensitivity.
FIXed: Fixes the internal level detector.
- AlcOnOffAutoEnum Mode#
// SCPI: [SOURce<HW>]:POWer:ALC:MODE AlcOnOffAutoEnum value = driver.Source.Power.Alc.Mode;Queries the currently set ALC mode. See method RsSmw.Source.Power.Alc.State.
powAlcMode: 0 | AUTO | 1| PRESet | OFFTable | ON | OFF | ONSample | ONTable
- AlcOffModeEnum Omode#
// SCPI: [SOURce<HW>]:POWer:ALC:OMODe AlcOffModeEnum value = driver.Source.Power.Alc.Omode; driver.Source.Power.Alc.Omode = AlcOffModeEnum.SHOLd;No help available.
- bool Search#
// SCPI: [SOURce<HW>]:POWer:ALC:SEARch bool value = driver.Source.Power.Alc.Search; driver.Source.Power.Alc.Search = false;No help available.
- PowAlcSampleLevEnum Slevel#
// SCPI: [SOURce<HW>]:POWer:ALC:SLEVel PowAlcSampleLevEnum value = driver.Source.Power.Alc.Slevel; driver.Source.Power.Alc.Slevel = PowAlcSampleLevEnum.ATTenuated;Sets the sample level of automatic level control (ALC) . How To: See “How to enable the ALC”.
sampLevel: FULL | MINimum | ATTenuated
- AlcOnOffAutoEnum State#
// SCPI: [SOURce<HW>]:POWer:ALC:[STATe] AlcOnOffAutoEnum value = driver.Source.Power.Alc.State; driver.Source.Power.Alc.State = AlcOnOffAutoEnum._0;Activates automatic level control in the selected mode. How to: See “How to enable the ALC”.
state: AUTO | OFFTable | ON | ONSample | ONTable | OFF
AUTO: Adjusts the output level to the operating conditions automatically.
OFFTable: Controls the level with the attenuation values of the internal ALC table.
ON: Activates internal level control permanently.
OFF: Deactivates internal level control, Sample & Hold mode is active.
ONSample: Starts the internal level control with the first change.
ONTable: Starts with the attenuation setting from the table and continues with automatic level control.
Subgroups