User
SCPI Command :
SENSe<CH>:[POWer]:FILTer:LENGth:[USER]
- class UserCls[source]
User commands group definition. 1 total commands, 0 Subgroups, 1 group commands
- get(channel=Channel.Default) float [source]
# SCPI: SENSe<CH>:[POWer]:FILTer:LENGth:[USER] value: float = driver.sense.power.filterPy.length.user.get(channel = repcap.Channel.Default)
Selects the filter length for SENS:POW:FILT:’TYPE USER. As the filter length works as a multiplier for the time window, a constant filter length results in a constant measurement time (see also ‘About the measuring principle, averaging filter, filter length, and achieving stable results’) .
INTRO_CMD_HELP: The R&S NRP power sensors provide different resolutions for setting the filter length, depending on the used sensor type:
Resolution = 1 for R&S NRPxx power sensors
Resolution = 2n for sensors of the R&S NRP-Zxx family, with n = 1 to 16
- param channel
optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Sense’)
- return
user: float Range: 1 to 65536
- set(user: float, channel=Channel.Default) None [source]
# SCPI: SENSe<CH>:[POWer]:FILTer:LENGth:[USER] driver.sense.power.filterPy.length.user.set(user = 1.0, channel = repcap.Channel.Default)
Selects the filter length for SENS:POW:FILT:’TYPE USER. As the filter length works as a multiplier for the time window, a constant filter length results in a constant measurement time (see also ‘About the measuring principle, averaging filter, filter length, and achieving stable results’) .
INTRO_CMD_HELP: The R&S NRP power sensors provide different resolutions for setting the filter length, depending on the used sensor type:
Resolution = 1 for R&S NRPxx power sensors
Resolution = 2n for sensors of the R&S NRP-Zxx family, with n = 1 to 16
- param user
float Range: 1 to 65536
- param channel
optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Sense’)