Dtime
SCPI Command :
[SENSe]:PMETer<p>:TRIGger:DTIMe
- Commands in total: 1Subgroups: 0Direct child commands: 1
- get(powerMeter=PowerMeter.Default) float[source]
# [SENSe]:PMETer<p>:TRIGger:DTIMe value: float = driver.sense.pmeter.trigger.dtime.get(powerMeter = repcap.PowerMeter.Default)
Defines the time period that the input signal has to stay below the IF power trigger level before the measurement starts.
- Parameters:
powerMeter – optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Pmeter’)
- Returns:
time: Range: 0 s to 1 s, Unit: S
- set(time: float, powerMeter=PowerMeter.Default) None[source]
# [SENSe]:PMETer<p>:TRIGger:DTIMe driver.sense.pmeter.trigger.dtime.set(time = 1.0, powerMeter = repcap.PowerMeter.Default)
Defines the time period that the input signal has to stay below the IF power trigger level before the measurement starts.
- Parameters:
time – Range: 0 s to 1 s, Unit: S
powerMeter – optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Pmeter’)