Hysteresis
SCPI Command :
[SENSe]:PMETer<p>:TRIGger:HYSTeresis
- class HysteresisCls[source]
Hysteresis commands group definition. 1 total commands, 0 Subgroups, 1 group commands
- get(powerMeter=PowerMeter.Default) float [source]
# SCPI: [SENSe]:PMETer<p>:TRIGger:HYSTeresis value: float = driver.sense.pmeter.trigger.hysteresis.get(powerMeter = repcap.PowerMeter.Default)
Defines the trigger hysteresis for external power triggers. The hysteresis in dB is the value the input signal must stay below the IF power trigger level to allow a trigger to start the measurement.
- param powerMeter
optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Pmeter’)
- return
hysteresis: Range: 3 dB to 50 dB, Unit: DB
- set(hysteresis: float, powerMeter=PowerMeter.Default) None [source]
# SCPI: [SENSe]:PMETer<p>:TRIGger:HYSTeresis driver.sense.pmeter.trigger.hysteresis.set(hysteresis = 1.0, powerMeter = repcap.PowerMeter.Default)
Defines the trigger hysteresis for external power triggers. The hysteresis in dB is the value the input signal must stay below the IF power trigger level to allow a trigger to start the measurement.
- param hysteresis
Range: 3 dB to 50 dB, Unit: DB
- param powerMeter
optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Pmeter’)