Hysteresis

SCPI Command :

[SENSe]:PMETer<p>:TRIGger:HYSTeresis
Commands in total: 1
Subgroups: 0
Direct child commands: 1
get(powerMeter=PowerMeter.Default) float[source]
# [SENSe]:PMETer<p>:TRIGger:HYSTeresis
value: float = driver.sense.pmeter.trigger.hysteresis.get(powerMeter = repcap.PowerMeter.Default)

Defines the trigger hysteresis for external power triggers. The hysteresis in dB is the value the input signal must stay below the IF power trigger level to allow a trigger to start the measurement.

Parameters:

powerMeter – optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Pmeter’)

Returns:

hysteresis: Range: 3 dB to 50 dB, Unit: DB

set(hysteresis: float, powerMeter=PowerMeter.Default) None[source]
# [SENSe]:PMETer<p>:TRIGger:HYSTeresis
driver.sense.pmeter.trigger.hysteresis.set(hysteresis = 1.0, powerMeter = repcap.PowerMeter.Default)

Defines the trigger hysteresis for external power triggers. The hysteresis in dB is the value the input signal must stay below the IF power trigger level to allow a trigger to start the measurement.

Parameters:
  • hysteresis – Range: 3 dB to 50 dB, Unit: DB

  • powerMeter – optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Pmeter’)