Achannel
SCPI Command :
[SENSe]:POWer:ACHannel:PRESet
- Commands in total: 64Subgroups: 15Direct child commands: 1
- preset(measurement: PowerMeasFunction) None[source]
# [SENSe]:POWer:ACHannel:PRESet driver.sense.power.achannel.preset(measurement = enums.PowerMeasFunction.ACPower)
Determines the ideal span, bandwidths and detector for the current power measurement. To get a valid result, you have to perform a complete measurement with synchronization to the end of the measurement before reading out the result. This is only possible for single sweep mode. See also method
RsFsw.applications.k10Xlte.initiate.continuous.set().- Parameters:
measurement – ACPower | MCACpower ACLR measurement CPOWer channel power measurement OBANdwidth | OBWidth Occupied bandwidth measurement CN Carrier to noise ratio CN0 Carrier to noise ration referenced to a 1 Hz bandwidth
Cloning the Group
# Create a copy of the original group, that exists independently
achannel_copy = driver.sense.power.achannel.clone()
Subgroups