Phold

SCPI Command :

CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:PHOLd
class PholdCls[source]

Phold commands group definition. 1 total commands, 0 Subgroups, 1 group commands

get(window=Window.Default, marker=Marker.Default) bool[source]
# SCPI: CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:PHOLd
value: bool = driver.calculate.marker.function.summary.phold.get(window = repcap.Window.Default, marker = repcap.Marker.Default)

Switches on or off the peak-hold function for the active power measurement in zero span in the window specified by the suffix <n>. If activated, the peak for each sweep is compared to the previously stored peak; the maximum of the two is stored as the current peak. The peak-hold function is reset by switching it off and on again.

param window

optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Calculate’)

param marker

optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Marker’)

return

state: ON | OFF | 1 | 0

set(state: bool, window=Window.Default, marker=Marker.Default) None[source]
# SCPI: CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:PHOLd
driver.calculate.marker.function.summary.phold.set(state = False, window = repcap.Window.Default, marker = repcap.Marker.Default)

Switches on or off the peak-hold function for the active power measurement in zero span in the window specified by the suffix <n>. If activated, the peak for each sweep is compared to the previously stored peak; the maximum of the two is stored as the current peak. The peak-hold function is reset by switching it off and on again.

param state

ON | OFF | 1 | 0

param window

optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Calculate’)

param marker

optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Marker’)