Average

SCPI Command :

CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:AVERage
Commands in total: 1
Subgroups: 0
Direct child commands: 1
get(window=Window.Default, marker=Marker.Default) bool[source]
# CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:AVERage
value: bool = driver.calculate.marker.function.summary.average.get(window = repcap.Window.Default, marker = repcap.Marker.Default)

Switches on or off averaging for the active power measurement in zero span in the window specified by the suffix <n>. If activated, a time domain value is calculated from the trace after each sweep; in the end, all values are averaged to calculate the final result. The number of results required for the calculation of average is defined with method RsFsw.applications.k10Xlte.sense.average.count.set() . Averaging is reset by switching it off and on again. Synchronization to the end of averaging is only possible in single sweep mode.

Parameters:
  • window – optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Calculate’)

  • marker – optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Marker’)

Returns:

state: ON | OFF | 1 | 0

set(state: bool, window=Window.Default, marker=Marker.Default) None[source]
# CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:AVERage
driver.calculate.marker.function.summary.average.set(state = False, window = repcap.Window.Default, marker = repcap.Marker.Default)

Switches on or off averaging for the active power measurement in zero span in the window specified by the suffix <n>. If activated, a time domain value is calculated from the trace after each sweep; in the end, all values are averaged to calculate the final result. The number of results required for the calculation of average is defined with method RsFsw.applications.k10Xlte.sense.average.count.set() . Averaging is reset by switching it off and on again. Synchronization to the end of averaging is only possible in single sweep mode.

Parameters:
  • state – ON | OFF | 1 | 0

  • window – optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Calculate’)

  • marker – optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Marker’)