IfPower

SCPI Command :

[SENSe]:SWEep:EGATe:LEVel:IFPower
Commands in total: 1
Subgroups: 0
Direct child commands: 1
get() float[source]
# [SENSe]:SWEep:EGATe:LEVel:IFPower
value: float = driver.applications.k14Xnr5G.sense.sweep.egate.level.ifPower.get()

Defines the power level at the third intermediate frequency that must be exceeded for the gate to be open. Note that any RF attenuation or preamplification is considered when the trigger level is analyzed. If defined, a reference level offset is also considered. Is only available for triggered gated measurements (method RsFsw.sense.sweep.egate.auto.set() MAN) .

set(level: float) None[source]
# [SENSe]:SWEep:EGATe:LEVel:IFPower
driver.applications.k14Xnr5G.sense.sweep.egate.level.ifPower.set(level = 1.0)

Defines the power level at the third intermediate frequency that must be exceeded for the gate to be open. Note that any RF attenuation or preamplification is considered when the trigger level is analyzed. If defined, a reference level offset is also considered. Is only available for triggered gated measurements (method RsFsw.sense.sweep.egate.auto.set() MAN) .

Parameters:

level – For details on available trigger levels and trigger bandwidths see the specifications document. Unit: DBM