IfPower

SCPI Command :

TRIGger[:SEQuence]:LEVel<ant>:IFPower
class IfPowerCls[source]

IfPower commands group definition. 1 total commands, 0 Subgroups, 1 group commands

get(instrument=Instrument.Default) float[source]
# SCPI: TRIGger[:SEQuence]:LEVel<ant>:IFPower
value: float = driver.applications.k10Xlte.trigger.sequence.level.ifPower.get(instrument = repcap.Instrument.Default)

Defines the power level at the third intermediate frequency that must be exceeded to cause a trigger event. Note that any RF attenuation or preamplification is considered when the trigger level is analyzed. If defined, a reference level offset is also considered.

param instrument

optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Level’)

return

level: numeric value For details on available trigger levels and trigger bandwidths see the specifications document. Unit: dBm

set(level: float, instrument=Instrument.Default) None[source]
# SCPI: TRIGger[:SEQuence]:LEVel<ant>:IFPower
driver.applications.k10Xlte.trigger.sequence.level.ifPower.set(level = 1.0, instrument = repcap.Instrument.Default)

Defines the power level at the third intermediate frequency that must be exceeded to cause a trigger event. Note that any RF attenuation or preamplification is considered when the trigger level is analyzed. If defined, a reference level offset is also considered.

param level

numeric value For details on available trigger levels and trigger bandwidths see the specifications document. Unit: dBm

param instrument

optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Level’)