Source:Frequency:Fixed#
SCPI Commands
- [SOURce<HW>]:FREQuency:[FIXed]:RCL
- [SOURce<HW>]:FREQuency:[FIXed]
Direct SCPI Members
- InclExclEnum Recall#
// SCPI: [SOURce<HW>]:FREQuency:[FIXed]:RCL InclExclEnum value = driver.Source.Frequency.Fixed.Recall; driver.Source.Frequency.Fixed.Recall = InclExclEnum.EXCLude;Set whether the RF frequency value is retained or taken from a loaded instrument configuration, when you recall instrument settings with command *RCL. The selected mode applies to both RF paths, i.e. a specified suffix is ignored.
rcl: INCLude | EXCLude
INCLude: Takes the frequency value of the loaded settings.
EXCLude: Retains the current frequency when an instrument configuration is loaded.
- double Value#
// SCPI: [SOURce<HW>]:FREQuency:[FIXed] double value = driver.Source.Frequency.Fixed.Value; driver.Source.Frequency.Fixed.Value = 1.0;Sets the frequency of the RF output signal in the selected path. The effect depends on the selected mode:
In CW mode (FREQ:MODE CW | FIXed) , the instrument operates at a fixed frequency.
In sweep mode (FREQ:MODE SWE) , the value applies to the sweep frequency. The instrument processes the frequency settings in defined sweep steps.
In user mode (FREQ:STEP:MODE USER) , you can vary the current frequency step by step.
For more information, refer to the specifications document.
fixedX: The following settings influence the value range: An offset set with the command method RsSmw.Source.Frequency.Offset()
Numerical value: Sets the frequency in CW and sweep mode
UP | DOWN: Varies the frequency step by step in user mode.The frequency is increased or decreased by the value set with the command [:SOURcehw]:FREQuency:STEP[:INCRement].