Source:Frequency:Fixed#

SCPI Commands

[SOURce<HW>]:FREQuency:[FIXed]:RCL
[SOURce<HW>]:FREQuency:[FIXed]

Direct SCPI Members

InclExclEnum Recall#
// SCPI: [SOURce<HW>]:FREQuency:[FIXed]:RCL
InclExclEnum value = driver.Source.Frequency.Fixed.Recall;
driver.Source.Frequency.Fixed.Recall = InclExclEnum.EXCLude;

Set whether the RF frequency value is retained or taken from a loaded instrument configuration, when you recall instrument settings with command *RCL. The selected mode applies to both RF paths, i.e. a specified suffix is ignored.

rcl: INCLude | EXCLude

  • INCLude: Takes the frequency value of the loaded settings.

  • EXCLude: Retains the current frequency when an instrument configuration is loaded.


double Value#
// SCPI: [SOURce<HW>]:FREQuency:[FIXed]
double value = driver.Source.Frequency.Fixed.Value;
driver.Source.Frequency.Fixed.Value = 1.0;

Sets the frequency of the RF output signal in the selected path. The effect depends on the selected mode:

  • In CW mode (FREQ:MODE CW | FIXed) , the instrument operates at a fixed frequency.

  • In sweep mode (FREQ:MODE SWE) , the value applies to the sweep frequency. The instrument processes the frequency settings in defined sweep steps.

  • In user mode (FREQ:STEP:MODE USER) , you can vary the current frequency step by step.

For more information, refer to the specifications document.

fixedX: The following settings influence the value range: An offset set with the command method RsSmw.Source.Frequency.Offset()

  • Numerical value: Sets the frequency in CW and sweep mode

  • UP | DOWN: Varies the frequency step by step in user mode.The frequency is increased or decreased by the value set with the command [:SOURcehw]:FREQuency:STEP[:INCRement].