Trigger
SCPI Commands :
[SOURce<HW>]:BB:NR5G:TRIGger:RMODe
[SOURce<HW>]:BB:NR5G:TRIGger:SLENgth
[SOURce<HW>]:BB:NR5G:TRIGger:SLUNit
[SOURce<HW>]:BB:NR5G:TRIGger:SOURce
[SOURce<HW>]:BB:NR5G:[TRIGger]:SEQuence
- class TriggerCls[source]
Trigger commands group definition. 44 total commands, 7 Subgroups, 5 group commands
- get_rmode() RsSmbv.enums.TrigRunMode [source]
# SCPI: [SOURce<HW>]:BB:NR5G:TRIGger:RMODe value: enums.TrigRunMode = driver.source.bb.nr5G.trigger.get_rmode()
Queries the signal generation status.
- return
trig_run_mode: STOP| RUN
- get_sequence() RsSmbv.enums.DmTrigMode [source]
# SCPI: [SOURce<HW>]:BB:NR5G:[TRIGger]:SEQuence value: enums.DmTrigMode = driver.source.bb.nr5G.trigger.get_sequence()
Sets the trigger mode.
- return
trig_mode: AUTO| RETRigger| AAUTo| ARETrigger| SINGle
- get_sl_unit() RsSmbv.enums.UnitSlB [source]
# SCPI: [SOURce<HW>]:BB:NR5G:TRIGger:SLUNit value: enums.UnitSlB = driver.source.bb.nr5G.trigger.get_sl_unit()
Defines the unit for the entry of the signal sequence length.
- return
trig_seq_len_unit: SEQuence| SAMPle
- get_slength() int [source]
# SCPI: [SOURce<HW>]:BB:NR5G:TRIGger:SLENgth value: int = driver.source.bb.nr5G.trigger.get_slength()
Defines the length of the signal sequence that is output in the SINGle trigger mode.
- return
trig_seq_len: integer Range: 1 to dynamic
- get_source() RsSmbv.enums.TriggerSourceC [source]
# SCPI: [SOURce<HW>]:BB:NR5G:TRIGger:SOURce value: enums.TriggerSourceC = driver.source.bb.nr5G.trigger.get_source() INTRO_CMD_HELP: Selects the trigger signal source and determines the way the triggering is executed. Provided are: - Internal triggering by a command (INTernal) - External trigger signal via one of the User x connectors EGT1: External global trigger - In primary-secondary instrument mode, the external baseband synchronization signal (BBSY) - EXTernal: Setting only Provided only for backward compatibility with other Rohde & Schwarz signal generators. The R&S SMBV100B accepts this value and maps it automatically as follows: EXTernal = EGT1 :return: trig_source: INTernal| EGT1| EXTernal| BBSY
- set_sequence(trig_mode: RsSmbv.enums.DmTrigMode) None [source]
# SCPI: [SOURce<HW>]:BB:NR5G:[TRIGger]:SEQuence driver.source.bb.nr5G.trigger.set_sequence(trig_mode = enums.DmTrigMode.AAUTo)
Sets the trigger mode.
- param trig_mode
AUTO| RETRigger| AAUTo| ARETrigger| SINGle
- set_sl_unit(trig_seq_len_unit: RsSmbv.enums.UnitSlB) None [source]
# SCPI: [SOURce<HW>]:BB:NR5G:TRIGger:SLUNit driver.source.bb.nr5G.trigger.set_sl_unit(trig_seq_len_unit = enums.UnitSlB.SAMPle)
Defines the unit for the entry of the signal sequence length.
- param trig_seq_len_unit
SEQuence| SAMPle
- set_slength(trig_seq_len: int) None [source]
# SCPI: [SOURce<HW>]:BB:NR5G:TRIGger:SLENgth driver.source.bb.nr5G.trigger.set_slength(trig_seq_len = 1)
Defines the length of the signal sequence that is output in the SINGle trigger mode.
- param trig_seq_len
integer Range: 1 to dynamic
- set_source(trig_source: RsSmbv.enums.TriggerSourceC) None [source]
# SCPI: [SOURce<HW>]:BB:NR5G:TRIGger:SOURce driver.source.bb.nr5G.trigger.set_source(trig_source = enums.TriggerSourceC.BBSY) INTRO_CMD_HELP: Selects the trigger signal source and determines the way the triggering is executed. Provided are: - Internal triggering by a command (INTernal) - External trigger signal via one of the User x connectors EGT1: External global trigger - In primary-secondary instrument mode, the external baseband synchronization signal (BBSY) - EXTernal: Setting only Provided only for backward compatibility with other Rohde & Schwarz signal generators. The R&S SMBV100B accepts this value and maps it automatically as follows: EXTernal = EGT1 :param trig_source: INTernal| EGT1| EXTernal| BBSY
Cloning the Group
# Create a clone of the original group, that exists independently
group2 = driver.source.bb.nr5G.trigger.clone()