Acquire

SCPI Command :

[SENSe]:CORRection:COLLect[:ACQuire]
class AcquireCls[source]

Acquire commands group definition. 1 total commands, 0 Subgroups, 1 group commands

set(meas_type: RsFsw.enums.CorrectionMeasType) None[source]
# SCPI: [SENSe]:CORRection:COLLect[:ACQuire]
driver.sense.correction.collect.acquire.set(meas_type = enums.CorrectionMeasType.OPEN)

Initiates a reference measurement (calibration) . The reference measurement is the basis for the measurement normalization. The result depends on whether a reflection measurement or transmission measurement is performed (see [SENSe:]CORRection:METHod) . To obtain a correct reference measurement, a complete sweep with synchronization to the end of the sweep must have been carried out. This is only possible in the single sweep mode. Is only available if External Generator Control (R&S FSW-B10) is installed and active.

param meas_type

THRough | OPEN THRough ‘TRANsmission’ mode: calibration with direct connection between generator and device input ‘REFLection’ mode: calibration with short circuit at the input OPEN only allowed in ‘REFLection’ mode: calibration with open input