Upper

SCPI Command :

[SENSe]:ADJust:CONFigure:HYSTeresis:UPPer
Commands in total: 1
Subgroups: 0
Direct child commands: 1
get() float[source]
# [SENSe]:ADJust:CONFigure:HYSTeresis:UPPer
value: float = driver.sense.adjust.configure.hysteresis.upper.get()

When the reference level is adjusted automatically using the method RsFsw.sense.adjust.level.set() command, the internal attenuators and the preamplifier are also adjusted. To avoid frequent adaptation due to small changes in the input signal, you can define a hysteresis. This setting defines an upper threshold the signal must exceed (compared to the last measurement) before the reference level is adapted automatically. This setting can only be adjusted in the MSRA primary application, not in the secondary applications.

Returns:

threshold: Range: 0 dB to 200 dB, Unit: dB

set(threshold: float) None[source]
# [SENSe]:ADJust:CONFigure:HYSTeresis:UPPer
driver.sense.adjust.configure.hysteresis.upper.set(threshold = 1.0)

When the reference level is adjusted automatically using the method RsFsw.sense.adjust.level.set() command, the internal attenuators and the preamplifier are also adjusted. To avoid frequent adaptation due to small changes in the input signal, you can define a hysteresis. This setting defines an upper threshold the signal must exceed (compared to the last measurement) before the reference level is adapted automatically. This setting can only be adjusted in the MSRA primary application, not in the secondary applications.

Parameters:

threshold – Range: 0 dB to 200 dB, Unit: dB