PreSelection

SCPI Command :

CALibration:PRESelection
class PreSelectionCls[source]

PreSelection commands group definition. 1 total commands, 0 Subgroups, 1 group commands

set() None[source]
# SCPI: CALibration:PRESelection
driver.calibration.preSelection.set()

Due to changes in temperature, the YIG-preselector frequency may become slightly offset. This command re-aligns the preselector quickly, without requiring a full self-alignment of the FSW. For FSW85 models, the YIG2 filter is also re-aligned. This command is only available for FSW models 1331.5003Kxx, and only if a YIG-preselector is available.

set_with_opc(opc_timeout_ms: int = - 1) None[source]
# SCPI: CALibration:PRESelection
driver.calibration.preSelection.set_with_opc()

Due to changes in temperature, the YIG-preselector frequency may become slightly offset. This command re-aligns the preselector quickly, without requiring a full self-alignment of the FSW. For FSW85 models, the YIG2 filter is also re-aligned. This command is only available for FSW models 1331.5003Kxx, and only if a YIG-preselector is available.

Same as set, but waits for the operation to complete before continuing further. Use the RsFsw.utilities.opc_timeout_set() to set the timeout value.

param opc_timeout_ms

Maximum time to wait in milliseconds, valid only for this call.