PreSelection
SCPI Command :
CALibration:PRESelection
- Commands in total: 1Subgroups: 0Direct child commands: 1
- set() None[source]
# CALibration:PRESelection driver.calibration.preSelection.set()
Due to changes in temperature, the YIG-preselector frequency may become slightly offset. This command re-aligns the preselector quickly, without requiring a full self-alignment of the FSW. For FSW85 models, the YIG2 filter is also re-aligned. This command is only available for FSW models 1331.5003Kxx, and only if a YIG-preselector is available.
- set_with_opc(opc_timeout_ms: int = -1) None[source]
# CALibration:PRESelection driver.calibration.preSelection.set_with_opc()
Due to changes in temperature, the YIG-preselector frequency may become slightly offset. This command re-aligns the preselector quickly, without requiring a full self-alignment of the FSW. For FSW85 models, the YIG2 filter is also re-aligned. This command is only available for FSW models 1331.5003Kxx, and only if a YIG-preselector is available.
Same as set, but waits for the operation to complete before continuing further. Use the RsFsw.utilities.opc_timeout_set() to set the timeout value.
- Parameters:
opc_timeout_ms – Maximum time to wait in milliseconds, valid only for this call.