Auto

SCPI Command :

CALCulate<n>:STATistics:SCALe:AUTO
Commands in total: 1
Subgroups: 0
Direct child commands: 1
set(event: EventOnce, window=Window.Default) None[source]
# CALCulate<n>:STATistics:SCALe:AUTO
driver.calculate.statistics.scale.auto.set(event = enums.EventOnce.ONCE, window = repcap.Window.Default)

Initiates an automatic scaling of the diagram (x- and y-axis) . To obtain maximum resolution, the level range is set as a function of the measured spacing between peak power and the minimum power for the APD measurement and of the spacing between peak power and mean power for the CCDF measurement. In addition, the probability scale for the number of test points is adapted. To get a valid result, you have to synchronize to the end of the auto range process in a complete sweep. This is only possible in single sweep mode. Note this command is not available when using an external frontend.

Parameters: