Result
SCPI Command :
CALCulate<n>:MARKer<m>:FUNCtion:NDBDown:RESult
- Commands in total: 1Subgroups: 0Direct child commands: 1
- get(window=Window.Default, marker=Marker.Default) float[source]
# CALCulate<n>:MARKer<m>:FUNCtion:NDBDown:RESult value: float = driver.calculate.marker.function.ndbDown.result.get(window = repcap.Window.Default, marker = repcap.Marker.Default)
Queries the distance of the n dB down markers from each other. To get a valid result, you have to perform a complete measurement with synchronization to the end of the measurement before reading out the result. This is only possible for single sweep mode. See also method
RsFsw.applications.k10Xlte.initiate.continuous.set().- Parameters:
window – optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Calculate’)
marker – optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Marker’)
- Returns:
distance: The result depends on the span. In case of frequency domain measurements, the command returns the bandwidth between the two n dB down markers in Hz. In case of time domain measurements, the command returns the pulse width between the two n dB down markers in seconds.