Result

SCPI Command :

CALCulate<n>:MARKer<m>:FUNCtion:NDBDown:RESult
class ResultCls[source]

Result commands group definition. 1 total commands, 0 Subgroups, 1 group commands

get(window=Window.Default, marker=Marker.Default) float[source]
# SCPI: CALCulate<n>:MARKer<m>:FUNCtion:NDBDown:RESult
value: float = driver.calculate.marker.function.ndbDown.result.get(window = repcap.Window.Default, marker = repcap.Marker.Default)

Queries the distance of the n dB down markers from each other. To get a valid result, you have to perform a complete measurement with synchronization to the end of the measurement before reading out the result. This is only possible for single sweep mode. See also method RsFsw.Applications.K10x_Lte.Initiate.Continuous.set.

param window

optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Calculate’)

param marker

optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Marker’)

return

distance: The result depends on the span. In case of frequency domain measurements, the command returns the bandwidth between the two n dB down markers in Hz. In case of time domain measurements, the command returns the pulse width between the two n dB down markers in seconds.