IfPower
SCPI Command :
TRIGger<tp>[:SEQuence]:LEVel:IFPower
- class IfPowerCls[source]
 IfPower commands group definition. 1 total commands, 0 Subgroups, 1 group commands
- get(triggerPort=TriggerPort.Default) float[source]
 # SCPI: TRIGger<tp>[:SEQuence]:LEVel:IFPower value: float = driver.applications.k70Vsa.trigger.sequence.level.ifPower.get(triggerPort = repcap.TriggerPort.Default)
Defines the power level at the third intermediate frequency that must be exceeded to cause a trigger event. Note that any RF attenuation or preamplification is considered when the trigger level is analyzed. If defined, a reference level offset is also considered.
- param triggerPort
 optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Trigger’)
- return
 level_if_power: No help available
- set(level_if_power: float, triggerPort=TriggerPort.Default) None[source]
 # SCPI: TRIGger<tp>[:SEQuence]:LEVel:IFPower driver.applications.k70Vsa.trigger.sequence.level.ifPower.set(level_if_power = 1.0, triggerPort = repcap.TriggerPort.Default)
Defines the power level at the third intermediate frequency that must be exceeded to cause a trigger event. Note that any RF attenuation or preamplification is considered when the trigger level is analyzed. If defined, a reference level offset is also considered.
- param level_if_power
 No help available
- param triggerPort
 optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Trigger’)