Holdoff
SCPI Command :
TRIGger<tp>[:SEQuence]:IFPower:HOLDoff
- class HoldoffCls[source]
Holdoff commands group definition. 1 total commands, 0 Subgroups, 1 group commands
- get(triggerPort=TriggerPort.Default) float [source]
# SCPI: TRIGger<tp>[:SEQuence]:IFPower:HOLDoff value: float = driver.applications.k70Vsa.trigger.sequence.ifPower.holdoff.get(triggerPort = repcap.TriggerPort.Default)
Defines the holding time before the next trigger event. Note that this command can be used for any trigger source, not just IF Power (despite the legacy keyword) . Note: If you perform gated measurements in combination with the IF Power trigger, the FSW ignores the holding time for frequency sweep, FFT sweep, zero span and I/Q data measurements.
- param triggerPort
optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Trigger’)
- return
period: Range: 0 s to 10 s, Unit: S
- set(period: float, triggerPort=TriggerPort.Default) None [source]
# SCPI: TRIGger<tp>[:SEQuence]:IFPower:HOLDoff driver.applications.k70Vsa.trigger.sequence.ifPower.holdoff.set(period = 1.0, triggerPort = repcap.TriggerPort.Default)
Defines the holding time before the next trigger event. Note that this command can be used for any trigger source, not just IF Power (despite the legacy keyword) . Note: If you perform gated measurements in combination with the IF Power trigger, the FSW ignores the holding time for frequency sweep, FFT sweep, zero span and I/Q data measurements.
- param period
Range: 0 s to 10 s, Unit: S
- param triggerPort
optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Trigger’)