End
SCPI Command :
CALCulate<n>:CHRDetection:POWer:OFFSet:END
- class EndCls[source]
End commands group definition. 1 total commands, 0 Subgroups, 1 group commands
- get(window=Window.Default) float [source]
# SCPI: CALCulate<n>:CHRDetection:POWer:OFFSet:END value: float = driver.applications.k60Transient.calculate.chrDetection.power.offset.end.get(window = repcap.Window.Default)
Defines the end of the measurement range for power results as an offset in seconds from the chirp end. This command is only available if the reference is EDGE (see method RsFsw.Applications.K60_Transient.Calculate.ChrDetection.Power. Reference.set) . For details on the measurement range parameters see ‘Measurement range’.
- param window
optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Calculate’)
- return
time: Unit: S
- set(time: float, window=Window.Default) None [source]
# SCPI: CALCulate<n>:CHRDetection:POWer:OFFSet:END driver.applications.k60Transient.calculate.chrDetection.power.offset.end.set(time = 1.0, window = repcap.Window.Default)
Defines the end of the measurement range for power results as an offset in seconds from the chirp end. This command is only available if the reference is EDGE (see method RsFsw.Applications.K60_Transient.Calculate.ChrDetection.Power. Reference.set) . For details on the measurement range parameters see ‘Measurement range’.
- param time
Unit: S
- param window
optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Calculate’)