Lower#
SCPI Command :
- [SENSe]:ADJust:CONFigure:HYSTeresis:LOWer
- Commands in total: 1Subgroups: 0Direct child commands: 1
- get ( ) float
- set ( threshold: float ) None
# [SENSe]:ADJust:CONFigure:HYSTeresis:LOWer # get: value: float = driver.applications.k14Xnr5G.sense.adjust.configure.hysteresis.lower.get() # set: driver.applications.k14Xnr5G.sense.adjust.configure.hysteresis.lower.set(threshold = 1.0)
When the reference level is adjusted automatically using the method
RsFsw.sense.adjust.level.set()command, the internal attenuators and the preamplifier are also adjusted. To avoid frequent adaptation due to small changes in the input signal, you can define a hysteresis. This setting defines a lower threshold the signal must fall below (compared to the last measurement) before the reference level is adapted automatically.- Parameters:
threshold – 0 dB to 200 dB