Lower
SCPI Command :
[SENSe]:ADJust:CONFigure:HYSTeresis:LOWer
- class LowerCls[source]
Lower commands group definition. 1 total commands, 0 Subgroups, 1 group commands
- get() float [source]
# SCPI: [SENSe]:ADJust:CONFigure:HYSTeresis:LOWer value: float = driver.applications.k10Xlte.sense.adjust.configure.hysteresis.lower.get()
When the reference level is adjusted automatically using the [SENSe:]ADJust:LEVel command, the internal attenuators and the preamplifier are also adjusted. To avoid frequent adaptation due to small changes in the input signal, you can define a hysteresis. This setting defines a lower threshold the signal must fall below (compared to the last measurement) before the reference level is adapted automatically. This setting can only be adjusted in the MSRA primary application, not in the secondary applications.
- return
threshold: Range: 0 dB to 200 dB, Unit: dB
- set(threshold: float) None [source]
# SCPI: [SENSe]:ADJust:CONFigure:HYSTeresis:LOWer driver.applications.k10Xlte.sense.adjust.configure.hysteresis.lower.set(threshold = 1.0)
When the reference level is adjusted automatically using the [SENSe:]ADJust:LEVel command, the internal attenuators and the preamplifier are also adjusted. To avoid frequent adaptation due to small changes in the input signal, you can define a hysteresis. This setting defines a lower threshold the signal must fall below (compared to the last measurement) before the reference level is adapted automatically. This setting can only be adjusted in the MSRA primary application, not in the secondary applications.
- param threshold
Range: 0 dB to 200 dB, Unit: dB