Trigger:Sweep:Source#

SCPI Commands

TRIGger<HW>:[SWEep]:SOURce

Direct SCPI Members

void Set ( SingExtAutoEnum source )#
void Set ( SingExtAutoEnum source, InputIxRepCap inputIx )
// SCPI: TRIGger<HW>:[SWEep]:SOURce
driver.Trigger.Sweep.Source.Set(SingExtAutoEnum.AUTO, InputIxRepCap.Default);
driver.Trigger.Sweep.Source.Set(SingExtAutoEnum.AUTO);

Selects the trigger source for the corresponding sweeps:

  • FSWeep - RF frequency

  • LFFSweep - LF frequency

  • PSWeep - RF level

  • SWEep - all sweeps

The source names of the parameters correspond to the values provided in manual control of the instrument. They differ from the SCPI-compliant names, but the instrument accepts both variants. Use the SCPI name, if compatibility is an important issue. Find the corresponding SCPI-compliant commands in .

source AUTO | IMMediate | SINGle | BUS | EXTernal | EAUTo

  • AUTO [IMMediate]: Executes a sweep automatically.In this free-running mode, the trigger condition is met continuously. I.e. when a sweep is completed, the next one starts immediately.

  • SINGle [BUS]: Executes one complete sweep cycle.The following commands initiate a trigger event:*TRG[:SOURcehw]:SWEep:POWer:EXECute[:SOURcehw]:SWEep[:FREQuency]:EXECute:TRIGgerhw[:SWEep][:IMMediate], :TRIGgerhw:PSWeep[:IMMediate] and :TRIGgerhw:FSWeep[:IMMediate].Set the sweep mode with the commands:[:SOURcehw]:SWEep:POWer:MODEAUTO | STEP[:SOURcehw]:SWEep[:FREQuency]:MODEAUTO | STEP[:SOURcehw]:LFOutput:SWEep[:FREQuency]:MODEAUTO | STEPIn step mode (STEP) , the instrument executes only one step.

  • EXTernal: An external signal triggers the sweep.

  • EAUTo: An external signal triggers the sweep. When one sweep is finished, the next sweep starts.A second trigger event stops the sweep at the current frequency, a third trigger event starts the trigger at the start frequency, and so on.

inputIx Repeated capability selector

When you do not use any explicit repeated capability selector, the value is taken from its parent group.