Source:Input:User:Trigger#
SCPI Commands
- [SOURce]:INPut:USER:TRIGger:IMPedance
- [SOURce]:INPut:USER:TRIGger:LEVel
- [SOURce]:INPut:USER:TRIGger:SLOPe
Direct SCPI Members
- ImpG50G1Kenum Impedance#
// SCPI: [SOURce]:INPut:USER:TRIGger:IMPedance ImpG50G1Kenum value = driver.Source.Input.User.Trigger.Impedance; driver.Source.Input.User.Trigger.Impedance = ImpG50G1Kenum.G1K;Selects the input impedance for the external trigger inputs.
impedance: G50 | G1K
G50: 50 Ω per ground
G1K: 1 kΩ per ground
- double Level#
// SCPI: [SOURce]:INPut:USER:TRIGger:LEVel double value = driver.Source.Input.User.Trigger.Level; driver.Source.Input.User.Trigger.Level = 1.0;Sets the threshold for any input signal at the USER 1 to USER 3 connectors.
level: float
- SlopeTypeEnum Slope#
// SCPI: [SOURce]:INPut:USER:TRIGger:SLOPe SlopeTypeEnum value = driver.Source.Input.User.Trigger.Slope; driver.Source.Input.User.Trigger.Slope = SlopeTypeEnum.NEGative;Sets the polarity of the active slope of an external trigger signal.
slope: NEGative | POSitive