Bler#
SCPI Commands
- BLER:LOAD
- BLER:PRESet
- BLER:RESult
- BLER:SEQuence
- BLER:STARt
- BLER:STATe
- BLER:STOP
- BLER:STORe
- BLER:UNIT
Cloning the Group
// Create a clone of the original group, that exists independently
var bler2 = driver.Bler.Clone();
Direct SCPI Members
- void Load ( string filename )#
// SCPI: BLER:LOAD driver.Bler.Load("abc");Loads the selected file from the default or the specified directory. Loaded are files with extension ber_bler Refer to for general information on file handling in the default and in a specific directory.
filenameFilename or complete file path; file extension can be omitted
- void Preset ( )#
- void PresetAndWait ( )#
- void PresetAndWait ( int opcTimeoutMs )
// SCPI: BLER:PRESet driver.Bler.Preset(); driver.Bler.PresetAndWait(); driver.Bler.PresetAndWait(1000);Sets the parameters of the BERT/BLER test generator to their default values (*RST values specified for the commands) . Not affected is the state set with the command .method RsSmw.Bert.State/method RsSmw.Bler.State.
PresetAndWait is the same as Preset, but waits for the operation to complete. The optional entered opcTimeoutMs is only valid for this call. If you omit the opcTimeoutMs, the method uses the settable value
RsSmw.Utilities.OpcTimeout.
- class Result_Data
Structure for reading output parameters.
- int NumBlock
Number of checked data bits.
- int NumErrors
Number of error bits.
- double ErrorRate
If no termination criterion has been reached since the beginning of the measurement, display the current quotient of NumErrors (number of error bits) and NumBlock (number of data bits) . When at least one final result has been reached in continuous measurement, display the most recent final result.
- int MeasFinished
Status of measurement.
1: Measurement has been terminated, i.e. stopped or the defined number of data bits or error bits has been reached.
0: Measurement has not been terminated.
- int ClockDetected
Status of clock line.
1: Clock line active.
0: Clock line not active.
- int DataDetected
Status of data line.
1: Data line active.Only clocked data is detected. If the clock signal is missing, a data change is also not detected.
0: Data line is not active.
- int Synchronized
integer
1: The measurement is synchronized and the ratio is assumed realistic.That is, the clock and data lines are active and the following apples:NumErrors/NumBlock 0.1.
0: The measurement is not synchronized.
- Result_Data Result#
// SCPI: BLER:RESult Result_Data value = driver.Bler.Result;Queries the result of the last BLER measurement and responds with seven results, separated by commas. The first measurement following the start also queries intermediate results for the number of data bits, error bits and error rate. For method RsSmw.Bler.Trigger.ModeAUTO, only the final results of each single measurement are queried in the following measurements. Note: The restart of a new measurement is delayed until the first measurement result has been queried. The resulting brief measurement interruption is irrelevant because the subsequent measurement is synchronized within 24 data bits.
For return value, see the help for Result_Data structure arguments
- BlerTrigModeEnum Sequence#
// SCPI: BLER:SEQuence BlerTrigModeEnum value = driver.Bler.Sequence; driver.Bler.Sequence = BlerTrigModeEnum.AUTO;No help available.
- void Start ( )#
- void StartAndWait ( )#
- void StartAndWait ( int opcTimeoutMs )
// SCPI: BLER:STARt driver.Bler.Start(); driver.Bler.StartAndWait(); driver.Bler.StartAndWait(1000);Starts a continuous measurement.
StartAndWait is the same as Start, but waits for the operation to complete. The optional entered opcTimeoutMs is only valid for this call. If you omit the opcTimeoutMs, the method uses the settable value
RsSmw.Utilities.OpcTimeout.
- bool State#
// SCPI: BLER:STATe bool value = driver.Bler.State; driver.Bler.State = false;Activates/deactivates the measurement.
state: 1 | ON | 0| OFF
- void Stop ( )#
- void StopAndWait ( )#
- void StopAndWait ( int opcTimeoutMs )
// SCPI: BLER:STOP driver.Bler.Stop(); driver.Bler.StopAndWait(); driver.Bler.StopAndWait(1000);Stops an ongoing measurement.
StopAndWait is the same as Stop, but waits for the operation to complete. The optional entered opcTimeoutMs is only valid for this call. If you omit the opcTimeoutMs, the method uses the settable value
RsSmw.Utilities.OpcTimeout.
- string Store#
// SCPI: BLER:STORe driver.Bler.Store = "abc";Saves the current settings into the selected file; the file extension (ber_bler) is assigned automatically. Refer to for general information on file handling in the default and in a specific directory.
Filename: Filename or complete file path.
- BertUnitEnum Unit#
// SCPI: BLER:UNIT BertUnitEnum value = driver.Bler.Unit; driver.Bler.Unit = BertUnitEnum.ENGineering;No help available.
Subgroups