Value
SCPI Command :
[SENSe]:ESPectrum<sb>:RANGe<ri>:INPut:GAIN[:VALue]
- class ValueCls[source]
Value commands group definition. 1 total commands, 0 Subgroups, 1 group commands
- get(subBlock=SubBlock.Default, rangePy=RangePy.Default) float [source]
# SCPI: [SENSe]:ESPectrum<sb>:RANGe<ri>:INPut:GAIN[:VALue] value: float = driver.sense.espectrum.range.inputPy.gain.value.get(subBlock = repcap.SubBlock.Default, rangePy = repcap.RangePy.Default)
This command selects the gain for a SEM range. In case of high speed measurements, the level of the preamplifier has to be identical for all ranges.
- param subBlock
optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Espectrum’)
- param rangePy
optional repeated capability selector. Default value: Ix1 (settable in the interface ‘Range’)
- return
gain: 15 dB | 30 dB The availability of preamplification levels depends on the FSW model. - FSW8/13: 15dB and 30 dB - FSW26 or higher: 30 dB All other values are rounded to the nearest of these two.
- set(gain: float, subBlock=SubBlock.Default, rangePy=RangePy.Default) None [source]
# SCPI: [SENSe]:ESPectrum<sb>:RANGe<ri>:INPut:GAIN[:VALue] driver.sense.espectrum.range.inputPy.gain.value.set(gain = 1.0, subBlock = repcap.SubBlock.Default, rangePy = repcap.RangePy.Default)
This command selects the gain for a SEM range. In case of high speed measurements, the level of the preamplifier has to be identical for all ranges.
- param gain
15 dB | 30 dB The availability of preamplification levels depends on the FSW model. - FSW8/13: 15dB and 30 dB - FSW26 or higher: 30 dB All other values are rounded to the nearest of these two.
- param subBlock
optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Espectrum’)
- param rangePy
optional repeated capability selector. Default value: Ix1 (settable in the interface ‘Range’)