Pspectrum

SCPI Command :

MMEMory:STORe<n>:PSPectrum
class PspectrumCls[source]

Pspectrum commands group definition. 1 total commands, 0 Subgroups, 1 group commands

get(store=Store.Default) str[source]
# SCPI: MMEMory:STORe<n>:PSPectrum
value: str = driver.massMemory.store.pspectrum.get(store = repcap.Store.Default)

Exports persistence spectrum data to an ASCii file. The file contains the most recently determined percentage value for each pixel in the persistence spectrum, that is, for 1001 frequency and 600 power values, followed by the 1001*600 maxhold percentages. For details see Table ‘ASCII file format for persistence spectrum trace export’. Note that, due to the large amount of data involved, the process of exporting the data can take a while. Secure User Mode In secure user mode, settings that are stored on the instrument are stored to volatile memory, which is restricted to 256 MB. Thus, a ‘memory limit reached’ error can occur although the hard disk indicates that storage space is still available. To store data permanently, select an external storage location such as a USB memory device. For details, see ‘Protecting Data Using the Secure User Mode’.

param store

optional repeated capability selector. Default value: Pos1 (settable in the interface ‘Store’)

return

filename: String containing the path and name of the target file.

set(filename: str, store=Store.Default) None[source]
# SCPI: MMEMory:STORe<n>:PSPectrum
driver.massMemory.store.pspectrum.set(filename = 'abc', store = repcap.Store.Default)

Exports persistence spectrum data to an ASCii file. The file contains the most recently determined percentage value for each pixel in the persistence spectrum, that is, for 1001 frequency and 600 power values, followed by the 1001*600 maxhold percentages. For details see Table ‘ASCII file format for persistence spectrum trace export’. Note that, due to the large amount of data involved, the process of exporting the data can take a while. Secure User Mode In secure user mode, settings that are stored on the instrument are stored to volatile memory, which is restricted to 256 MB. Thus, a ‘memory limit reached’ error can occur although the hard disk indicates that storage space is still available. To store data permanently, select an external storage location such as a USB memory device. For details, see ‘Protecting Data Using the Secure User Mode’.

param filename

String containing the path and name of the target file.

param store

optional repeated capability selector. Default value: Pos1 (settable in the interface ‘Store’)