Table

SCPI Command :

MMEMory:STORe<n>:TABLe
class TableCls[source]

Table commands group definition. 1 total commands, 0 Subgroups, 1 group commands

set(columns: RsFsw.enums.StatisticType, filename: str, store=Store.Default) None[source]
# SCPI: MMEMory:STORe<n>:TABLe
driver.applications.k60Transient.massMemory.store.table.set(columns = enums.StatisticType.ALL, filename = 'abc', store = repcap.Store.Default)

Exports result table data from the specified window to an ASCii file (.DAT) . For details on the file format see ‘Reference: ASCII file export format’. Secure User Mode In secure user mode, settings that are stored on the instrument are stored to volatile memory, which is restricted to 256 MB. Thus, a ‘memory limit reached’ error can occur although the hard disk indicates that storage space is still available. To store data permanently, select an external storage location such as a USB memory device. For details, see ‘Protecting Data Using the Secure User Mode’.

param columns

Columns to be stored in file SELected Export only the selected (visible) table columns ALL Export all table columns (all possible measured parameters)

param filename

String containing the path and name of the target file.

param store

optional repeated capability selector. Default value: Pos1 (settable in the interface ‘Store’)