Pmode

SCPI Command :

[SENSe]:PROBe<pb>:SETup:PMODe
class PmodeCls[source]

Pmode commands group definition. 1 total commands, 0 Subgroups, 1 group commands

get(probe=Probe.Default) RsFsw.enums.ProbeMode[source]
# SCPI: [SENSe]:PROBe<pb>:SETup:PMODe
value: enums.ProbeMode = driver.applications.k18AmplifierEt.sense.probe.setup.pmode.get(probe = repcap.Probe.Default)

Determines the mode of a multi-mode modular probe. For details see ‘Using Probes’.

param probe

optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Probe’)

return

mode: CM | DM | PM | NM DM Voltage between the positive and negative input terminal CM Mean voltage between the positive and negative input terminal vs. ground PM Voltage between the positive input terminal and ground NM Voltage between the negative input terminal and ground

set(mode: RsFsw.enums.ProbeMode, probe=Probe.Default) None[source]
# SCPI: [SENSe]:PROBe<pb>:SETup:PMODe
driver.applications.k18AmplifierEt.sense.probe.setup.pmode.set(mode = enums.ProbeMode.CM, probe = repcap.Probe.Default)

Determines the mode of a multi-mode modular probe. For details see ‘Using Probes’.

param mode

CM | DM | PM | NM DM Voltage between the positive and negative input terminal CM Mean voltage between the positive and negative input terminal vs. ground PM Voltage between the positive input terminal and ground NM Voltage between the negative input terminal and ground

param probe

optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Probe’)