Pmode#

SCPI Command :

[SENSe]:PROBe<pb>:SETup:PMODe
Commands in total: 1
Subgroups: 0
Direct child commands: 1
get ( probe=Probe.Default ) → ProbeMode
set ( mode: ProbeMode, probe=Probe.Default ) → None
# [SENSe]:PROBe<pb>:SETup:PMODe
# get:
value: enums.ProbeMode = driver.applications.k17Mcgd.sense.probe.setup.pmode.get(probe = repcap.Probe.Default)
# set:
driver.applications.k17Mcgd.sense.probe.setup.pmode.set(mode = enums.ProbeMode.CM, probe = repcap.Probe.Default)

Determines the mode of a multi-mode modular probe. For details see .

Parameters:
  • mode – CM | DM | PM | NM

    • DM: Voltage between the positive and negative input terminal

    • CM: Mean voltage between the positive and negative input terminal vs. ground

    • PM: Voltage between the positive input terminal and ground

    • NM: Voltage between the negative input terminal and ground

  • probe – optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Probe’)