NmOffset#
SCPI Command :
- [SENSe]:PROBe<pb>:SETup:NMOFfset
- Commands in total: 1Subgroups: 0Direct child commands: 1
- get ( probe=Probe.Default ) float
- set ( nm_offset: float, probe=Probe.Default ) None
# [SENSe]:PROBe<pb>:SETup:NMOFfset # get: value: float = driver.applications.k17Mcgd.sense.probe.setup.nmOffset.get(probe = repcap.Probe.Default) # set: driver.applications.k17Mcgd.sense.probe.setup.nmOffset.set(nm_offset = 1.0, probe = repcap.Probe.Default)
Sets the N-mode offset. The setting is only available if a modular probe in N-mode is connected to the FSW. The maximum voltage difference between the positive and negative input terminals is 16 V. If the probe is disconnected, the N-mode offset of the probe is reset to 0.0 V. Note that if the offset for DM-mode or CM-mode is changed, the offsets for the P-mode and N-mode are adapted accordingly, and vice versa. For details see .
- Parameters:
nm_offset – The voltage offset between the negative input terminal and ground.
probe – optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Probe’)