Select#
SCPI Command :
- INPut:SELect
- Commands in total: 1Subgroups: 0Direct child commands: 1
- get ( ) BbInputSource
- set ( signal_source: BbInputSource ) None
# INPut:SELect # get: value: enums.BbInputSource = driver.applications.k17Mcgd.inputPy.select.get() # set: driver.applications.k17Mcgd.inputPy.select.set(signal_source = enums.BbInputSource.AIQ)
Selects the signal source for measurements, i.e. it defines which connector is used to input data to the FSW. If no additional input options are installed, only RF input or file input is supported. For FSW85 models with two RF input connectors, you must select the input connector to configure first using method
RsFsw.applications.k10Xlte.inputPy.typePy.set(). For Multicarrier Wideband Noise (MCWN) measurements, only RF input is allowed. Tip: The I/Q data to be analyzed for GSM cannot only be measured by the R&S FSW GSM application itself, it can also be imported to the application, provided it has the correct format. Furthermore, the analyzed I/Q data from the R&S FSW GSM application can be exported for further analysis in external applications. For details, see the FSW I/Q Analyzer and I/Q Input User Manual.- Parameters:
signal_source –
RF: Radio Frequency (RF INPUT connector)
FIQ: I/Q data fileFor details, see ‘Basics on Input from I/Q Data Files’.Not available for Input2.
DIQ: Digital IQ data (only available with optional Digital Baseband interface) For details on I/Q input see ‘Digital Input’.Not available for Input2.
AIQ: Analog Baseband signal (only available with optional Analog Baseband interface) For details on Analog Baseband input, see ‘Processing Data From the Analog Baseband Interface’.Not available for Input2.