Measurement

SCPI Command :

CONFigure[:NR5G]:MEASurement
class MeasurementCls[source]

Measurement commands group definition. 1 total commands, 0 Subgroups, 1 group commands

get() RsFsw.enums.MeasurementNr5G[source]
# SCPI: CONFigure[:NR5G]:MEASurement
value: enums.MeasurementNr5G = driver.applications.k14Xnr5G.configure.nr5G.measurement.get()

Selects the measurement type.

return

measurement: ACLR Selects the adjacent channel leakage ratio (ACLR) measurement. CACLr Selects the Cumulative ACLR measurement. CMEasurement Selects combined EVM / ACLR / SEM measurement. ESPectrum Selects the spectrum emission mask (SEM) measurement. EVM Selects I/Q measurements. MCAClr Selects Multi-Carrier ACLR measurement. MCESpectrum Selects Multi-Carrier SEM measurement. TAERor Selects the time alignment error measurement. TPOO Selects the transmit on / off power measurement.

set(measurement: RsFsw.enums.MeasurementNr5G) None[source]
# SCPI: CONFigure[:NR5G]:MEASurement
driver.applications.k14Xnr5G.configure.nr5G.measurement.set(measurement = enums.MeasurementNr5G.ACLR)

Selects the measurement type.

param measurement

ACLR Selects the adjacent channel leakage ratio (ACLR) measurement. CACLr Selects the Cumulative ACLR measurement. CMEasurement Selects combined EVM / ACLR / SEM measurement. ESPectrum Selects the spectrum emission mask (SEM) measurement. EVM Selects I/Q measurements. MCAClr Selects Multi-Carrier ACLR measurement. MCESpectrum Selects Multi-Carrier SEM measurement. TAERor Selects the time alignment error measurement. TPOO Selects the transmit on / off power measurement.