Resolution
SCPI Command :
CALCulate<n>:MARKer<m>:COUNt:RESolution
- Commands in total: 1Subgroups: 0Direct child commands: 1
- get(window=Window.Default, marker=Marker.Default) float[source]
# CALCulate<n>:MARKer<m>:COUNt:RESolution value: float = driver.applications.k14Xnr5G.calculate.marker.count.resolution.get(window = repcap.Window.Default, marker = repcap.Marker.Default)
Defines the resolution of the frequency counter.
- Parameters:
window – optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Calculate’)
marker – optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Marker’)
- Returns:
resolution: 0.001 | 0.01 | 0.1 | 1 | 10 | 100 | 1000 | 10000 Hz Unit: HZ
- set(resolution: float, window=Window.Default, marker=Marker.Default) None[source]
# CALCulate<n>:MARKer<m>:COUNt:RESolution driver.applications.k14Xnr5G.calculate.marker.count.resolution.set(resolution = 1.0, window = repcap.Window.Default, marker = repcap.Marker.Default)
Defines the resolution of the frequency counter.
- Parameters:
resolution – 0.001 | 0.01 | 0.1 | 1 | 10 | 100 | 1000 | 10000 Hz Unit: HZ
window – optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Calculate’)
marker – optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Marker’)