RfPower
SCPI Command :
[SENSe]:SWEep:EGATe:LEVel<ant>:RFPower
- Commands in total: 1Subgroups: 0Direct child commands: 1
- get(instrument=Instrument.Default) float[source]
# [SENSe]:SWEep:EGATe:LEVel<ant>:RFPower value: float = driver.applications.k10Xlte.sense.sweep.egate.level.rfPower.get(instrument = repcap.Instrument.Default)
Defines the gate level for which the gate is open. Note that any RF attenuation or preamplification is considered when the trigger level is analyzed. If defined, a reference level offset is also considered. The input signal must be between 500 MHz and 8 GHz. Is only available for triggered gated measurements (method
RsFsw.sense.sweep.egate.auto.set()MAN) .- Parameters:
instrument – optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Level’)
- set(level: float, instrument=Instrument.Default) None[source]
# [SENSe]:SWEep:EGATe:LEVel<ant>:RFPower driver.applications.k10Xlte.sense.sweep.egate.level.rfPower.set(level = 1.0, instrument = repcap.Instrument.Default)
Defines the gate level for which the gate is open. Note that any RF attenuation or preamplification is considered when the trigger level is analyzed. If defined, a reference level offset is also considered. The input signal must be between 500 MHz and 8 GHz. Is only available for triggered gated measurements (method
RsFsw.sense.sweep.egate.auto.set()MAN) .- Parameters:
level – For details on available trigger levels and trigger bandwidths, refer to the specifications document. Unit: DBM
instrument – optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Level’)