RfPower

SCPI Command :

[SENSe]:SWEep:EGATe:LEVel<ant>:RFPower
class RfPowerCls[source]

RfPower commands group definition. 1 total commands, 0 Subgroups, 1 group commands

get(instrument=Instrument.Default) float[source]
# SCPI: [SENSe]:SWEep:EGATe:LEVel<ant>:RFPower
value: float = driver.applications.k10Xlte.sense.sweep.egate.level.rfPower.get(instrument = repcap.Instrument.Default)

Defines the gate level for which the gate is open. Note that any RF attenuation or preamplification is considered when the trigger level is analyzed. If defined, a reference level offset is also considered. The input signal must be between 500 MHz and 8 GHz. Is only available for triggered gated measurements ([SENSe:]SWEep:EGATe:AUTOMAN) .

param instrument

optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Level’)

return

level: No help available

set(level: float, instrument=Instrument.Default) None[source]
# SCPI: [SENSe]:SWEep:EGATe:LEVel<ant>:RFPower
driver.applications.k10Xlte.sense.sweep.egate.level.rfPower.set(level = 1.0, instrument = repcap.Instrument.Default)

Defines the gate level for which the gate is open. Note that any RF attenuation or preamplification is considered when the trigger level is analyzed. If defined, a reference level offset is also considered. The input signal must be between 500 MHz and 8 GHz. Is only available for triggered gated measurements ([SENSe:]SWEep:EGATe:AUTOMAN) .

param level

For details on available trigger levels and trigger bandwidths, refer to the specifications document. Unit: DBM

param instrument

optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Level’)