Count
SCPI Command :
[SENSe]:PMETer<p>:MTIMe:AVERage:COUNt
- class CountCls[source]
Count commands group definition. 1 total commands, 0 Subgroups, 1 group commands
- get(powerMeter=PowerMeter.Default) float [source]
# SCPI: [SENSe]:PMETer<p>:MTIMe:AVERage:COUNt value: float = driver.applications.k10Xlte.sense.pmeter.mtime.average.count.get(powerMeter = repcap.PowerMeter.Default)
Sets the number of power readings included in the averaging process of power sensor measurements. Extended averaging yields more stable results for power sensor measurements, especially for measurements on signals with a low power, because it minimizes the effects of noise.
- param powerMeter
optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Pmeter’)
- return
number_readings: An average count of 0 or 1 performs one power reading. Range: 0 to 256
- set(number_readings: float, powerMeter=PowerMeter.Default) None [source]
# SCPI: [SENSe]:PMETer<p>:MTIMe:AVERage:COUNt driver.applications.k10Xlte.sense.pmeter.mtime.average.count.set(number_readings = 1.0, powerMeter = repcap.PowerMeter.Default)
Sets the number of power readings included in the averaging process of power sensor measurements. Extended averaging yields more stable results for power sensor measurements, especially for measurements on signals with a low power, because it minimizes the effects of noise.
- param number_readings
An average count of 0 or 1 performs one power reading. Range: 0 to 256
- param powerMeter
optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Pmeter’)