State
SCPI Command :
TRACe:IQ[:STATe]
- class StateCls[source]
State commands group definition. 1 total commands, 0 Subgroups, 1 group commands
- get() bool [source]
# SCPI: TRACe:IQ[:STATe] value: bool = driver.applications.iqAnalyzer.trace.iq.state.get()
Activates the simple I/Q data acquisition mode (see ‘Different remote modes available’) . Executing this command also has the following effects:
INTRO_CMD_HELP: For triggered gated measurements,only the following gate trigger sources are supported:
The sweep, amplitude, input and trigger settings from the measurement are retained.
All measurements are turned off.
All traces are set to ‘Blank’ mode.
The I/Q data analysis mode is turned off (TRAC:IQ:EVAL OFF) .
Note: To turn trace display back on or to enable the evaluation functions of the I/Q Analyzer, execute the TRAC:IQ:EVAL ON command (see method RsFsw.Trace.Iq.Eval.set) .
- return
state: ON | OFF | 0 | 1 OFF | 0 Switches the function off ON | 1 Switches the function on
- set(state: bool) None [source]
# SCPI: TRACe:IQ[:STATe] driver.applications.iqAnalyzer.trace.iq.state.set(state = False)
Activates the simple I/Q data acquisition mode (see ‘Different remote modes available’) . Executing this command also has the following effects:
INTRO_CMD_HELP: For triggered gated measurements,only the following gate trigger sources are supported:
The sweep, amplitude, input and trigger settings from the measurement are retained.
All measurements are turned off.
All traces are set to ‘Blank’ mode.
The I/Q data analysis mode is turned off (TRAC:IQ:EVAL OFF) .
Note: To turn trace display back on or to enable the evaluation functions of the I/Q Analyzer, execute the TRAC:IQ:EVAL ON command (see method RsFsw.Trace.Iq.Eval.set) .
- param state
ON | OFF | 0 | 1 OFF | 0 Switches the function off ON | 1 Switches the function on