Source:Roscillator:Internal:Adjust#

SCPI Commands

[SOURce]:ROSCillator:[INTernal]:ADJust:VALue
[SOURce]:ROSCillator:[INTernal]:ADJust:[STATe]

Direct SCPI Members

int Value#
// SCPI: [SOURce]:ROSCillator:[INTernal]:ADJust:VALue
int value = driver.Source.Roscillator.Internal.Adjust.Value;
driver.Source.Roscillator.Internal.Adjust.Value = 1;

Specifies the frequency correction value (adjustment value) .

value: integer


bool State#
// SCPI: [SOURce]:ROSCillator:[INTernal]:ADJust:[STATe]
bool value = driver.Source.Roscillator.Internal.Adjust.State;
driver.Source.Roscillator.Internal.Adjust.State = false;

Determines whether the calibrated (off) or a user-defined (on) adjustment value is used for fine adjustment of the frequency.

state: 1 | ON | 0| OFF

  • 0: Fine adjustment with the calibrated frequency value

  • 1: User-defined adjustment value.The instrument is no longer in the calibrated state.The calibration value is, however, not changed. The instrument resumes the calibrated state if you send SOURce:ROSCillator:INTernal:ADJust:STATe 0.