Source:List:Trigger#

SCPI Commands

[SOURce<HW>]:LIST:TRIGger:SOURce

Cloning the Group

// Create a clone of the original group, that exists independently
var trigger2 = driver.Source.List.Trigger.Clone();

Direct SCPI Members

TrigSweepSourNoHopExtAutoEnum Source#
// SCPI: [SOURce<HW>]:LIST:TRIGger:SOURce
TrigSweepSourNoHopExtAutoEnum value = driver.Source.List.Trigger.Source;
driver.Source.List.Trigger.Source = TrigSweepSourNoHopExtAutoEnum.AUTO;

Selects the trigger source for processing lists. The designation of the parameters correspond to those in sweep mode. SCPI standard uses other designations for the parameters, which are also accepted by the instrument. The SCPI designation should be used if compatibility is an important consideration. For an overview, see the following table: Rohde & Schwarz parameter / SCPI parameter / Applies to the list mode parameters:

  • AUTO / IMMediate / [:SOURce<hw>]:LIST:MODE AUTO

  • SINGle / BUS / [:SOURce<hw>]:LIST:MODE AUTO or [:SOURce<hw>]:LIST:MODE STEP

  • EXTernal / EXTernal / [:SOURce<hw>]:LIST:MODE AUTO or [:SOURce<hw>]:LIST:MODE STEP

source: AUTO | IMMediate | SINGle | BUS | EXTernal

  • AUTO | IMMediate: The trigger is free-running, i.e. the trigger condition is fulfilled continuously. The selected list is restarted as soon as it is finished.

  • SINGle | BUS: The list is triggered by the command [:SOURcehw]:LIST:TRIGger:EXECute. The list is executed once.

  • EXTernal: The list is triggered externally and executed once.