Source:LfOutput:Frequency#

SCPI Commands

[SOURce<HW>]:LFOutput<CH>:FREQuency
[SOURce<HW>]:LFOutput:FREQuency:MANual
[SOURce<HW>]:LFOutput:FREQuency:MODE
[SOURce<HW>]:LFOutput:FREQuency:STARt
[SOURce<HW>]:LFOutput:FREQuency:STOP

Direct SCPI Members

double Get ( )#
double Get ( LfOutputRepCap lfOutput )
void Set ( double frequency )#
void Set ( double frequency, LfOutputRepCap lfOutput )
// SCPI: [SOURce<HW>]:LFOutput<CH>:FREQuency
// Get:
double value = driver.Source.LfOutput.Frequency.Get();
double value = driver.Source.LfOutput.Frequency.Get(LfOutputRepCap.Default);
// Set:
driver.Source.LfOutput.Frequency.Set(1.0, LfOutputRepCap.Default);
driver.Source.LfOutput.Frequency.Set(1.0);

Sets the frequency of the LF signal in method RsSmw.Source.LfOutput.Frequency.Mode CW | FIXed mode. Note:

  • If the LF generator is used as a signal source, the instrument performs the analog modulations (AM/FM/ΦM/PM) with this frequency.

  • In sweep mode (method RsSmw.Source.LfOutput.Frequency.Mode SWE) , the frequency is coupled with the sweep frequency.

frequency float

lfOutput Repeated capability selector

When you do not use any explicit repeated capability selector, the value is taken from its parent group.


double Manual#
// SCPI: [SOURce<HW>]:LFOutput:FREQuency:MANual
double value = driver.Source.LfOutput.Frequency.Manual;
driver.Source.LfOutput.Frequency.Manual = 1.0;

Sets the frequency of the subsequent sweep step if LFO:SWE:MODE MAN. Use a separate command for each sweep step.

manual: You can select any value within the setting range, where: STARt is set with method RsSmw.Source.LfOutput.Frequency.Start() STOP is set with method RsSmw.Source.LfOutput.Frequency.Stop()


LfFreqModeEnum Mode#
// SCPI: [SOURce<HW>]:LFOutput:FREQuency:MODE
LfFreqModeEnum value = driver.Source.LfOutput.Frequency.Mode;
driver.Source.LfOutput.Frequency.Mode = LfFreqModeEnum.CW;

Sets the mode for the output of the LF generator frequency, and determines the commands to be used for frequency settings.

mode: CW | FIXed | SWEep

  • CW | FIXed: Sets the fixed-frequency mode. CW and FIXed are synonyms.To set the output frequency, use command [:SOURcehw]:LFOutputch:FREQuency

  • SWEep: Sets sweep mode.To set the frequency, use the commands:[:SOURcehw]:LFOutput:FREQuency:STARt and [:SOURcehw]:LFOutput:FREQuency:STOPOr [:SOURcehw]:LFOutput:FREQuency:MANual


double Start#
// SCPI: [SOURce<HW>]:LFOutput:FREQuency:STARt
double value = driver.Source.LfOutput.Frequency.Start;
driver.Source.LfOutput.Frequency.Start = 1.0;

Sets the start/stop frequency for method RsSmw.Source.LfOutput.Frequency.Mode SWEep.

start: float


double Stop#
// SCPI: [SOURce<HW>]:LFOutput:FREQuency:STOP
double value = driver.Source.LfOutput.Frequency.Stop;
driver.Source.LfOutput.Frequency.Stop = 1.0;

Sets the start/stop frequency for method RsSmw.Source.LfOutput.Frequency.Mode SWEep.