Source:Efrontend#
SCPI Commands
- [SOURce<HW>]:EFRontend:AMODe
- [SOURce<HW>]:EFRontend:CMODe
- [SOURce<HW>]:EFRontend:IDN
- [SOURce<HW>]:EFRontend:INFO
- [SOURce<HW>]:EFRontend:IPADdress
- [SOURce<HW>]:EFRontend:LIST
- [SOURce<HW>]:EFRontend:OPT
- [SOURce<HW>]:EFRontend:RFConnector
Cloning the Group
// Create a clone of the original group, that exists independently
var efrontend2 = driver.Source.Efrontend.Clone();
Direct SCPI Members
- EfePowAttModeEnum Amode#
// SCPI: [SOURce<HW>]:EFRontend:AMODe EfePowAttModeEnum value = driver.Source.Efrontend.Amode; driver.Source.Efrontend.Amode = EfePowAttModeEnum.AOFFset;Sets the attenuator mode of the external frontend.
attenuationMode: AUTO | MANual | AOFFset
AUTO: Sets the attenuation value automatically to the attenuation value provided from the connected external frontend.
MANual: Sets an attenuation value manually.
AOFFset: Requires frontend R&S FE170ST or R&S FE110ST.You can define an offset value which is added to the attenuation value provided from the connected external frontend.
- ConModeEnum Cmode#
// SCPI: [SOURce<HW>]:EFRontend:CMODe ConModeEnum value = driver.Source.Efrontend.Cmode; driver.Source.Efrontend.Cmode = ConModeEnum.AUTO;Sets the mode of the SSL control connection between R&S SMW200A and external frontend.
feConnMode: AUTO | LOCK | RXTX
AUTO: The R&S SMW200A locks external frontend, when activating the RF output at the R&S SMW200A (:OUTPut1 ON ) for output of the IF signal.The R&S SMW200A unlocks external frontend, when deactivating the RF output at the R&S SMW200A (:OUTPut1 OFF) .
LOCK: The external frontend is locked permanently. No other instrument can take over control.
RXTX: Requires an R&S FE50DTR connected to the R&S SMW200A.Connection mode for simultaneous Rx (receive) operation and Tx (transmit) operation of an R&S FE50DTR.Set this mode, if your test setup requires a release of connection lock. A connected vector signal analyzer can lock the external frontend for Rx operation.
- string Idn#
// SCPI: [SOURce<HW>]:EFRontend:IDN string value = driver.Source.Efrontend.Idn;Identification Returns the IDN string, i.e. the identification of the external frontend.
idnString: string
- string Info#
// SCPI: [SOURce<HW>]:EFRontend:INFO string value = driver.Source.Efrontend.Info;Queries information about the connected external frontend.
feInfo: string
- string IpAddress#
// SCPI: [SOURce<HW>]:EFRontend:IPADdress string value = driver.Source.Efrontend.IpAddress;Queries the IP address of the connected external frontend.
ipAddress: string
- List<string>List#
// SCPI: [SOURce<HW>]:EFRontend:LIST List<string> value = driver.Source.Efrontend.List;Queries connected external frontends in a comma-separated list.
filter: String
- string Opt#
// SCPI: [SOURce<HW>]:EFRontend:OPT string value = driver.Source.Efrontend.Opt;Option identification query Queries the options included in the external frontend. For more information, refer to the specifications document.
optString: The query returns a list of options. The options are returned at fixed positions in a comma-separated string. A zero is returned for options that are not installed.
- FenUmbRfConEnum RfConnector#
// SCPI: [SOURce<HW>]:EFRontend:RFConnector FenUmbRfConEnum value = driver.Source.Efrontend.RfConnector; driver.Source.Efrontend.RfConnector = FenUmbRfConEnum.NONE;Queries the active RF output connector at the connected RF frontend.
feOutputPath: NONE | RFA | RFB
NONE: No frontend connected.
RFA: Output connector RF A is active at the external frontend.
RFB: Output connector RF B is active at the external frontend.
Subgroups
- Source:Efrontend:Alignment
- Source:Efrontend:Cal
- Source:Efrontend:Concurrent
- Source:Efrontend:Connection
- Source:Efrontend:ExtDevice<ExternalDeviceRepCap>
- Source:Efrontend:FirmwareUpdate
- Source:Efrontend:Frequency
- Source:Efrontend:Loscillator
- Source:Efrontend:Network
- Source:Efrontend:Power
- Source:Efrontend:Screw
- Source:Efrontend:Selftest
- Source:Efrontend:TrxMode