Calibration:Delay#
SCPI Commands
- CALibration:DELay:MINutes
- CALibration:DELay:[MEASure]
Cloning the Group
// Create a clone of the original group, that exists independently
var delay2 = driver.Calibration.Delay.Clone();
Direct SCPI Members
- int Minutes#
// SCPI: CALibration:DELay:MINutes int value = driver.Calibration.Delay.Minutes; driver.Calibration.Delay.Minutes = 1;Sets the warm-up time to wait before internal adjustment starts automatically. Automatic execution starts only, if you have enabled the calibration with command ON.
minutes: integer
- bool Measure#
// SCPI: CALibration:DELay:[MEASure] bool value = driver.Calibration.Delay.Measure;Starts the delayed adjustment process. When the warm-up time has elapsed (see , it executes the internal adjustments. If you have enabled automatic shutdown, , the instrument shuts down when the adjustments are completed.
error: 1 | ON | 0| OFF
Subgroups