Bert:Setup#

SCPI Commands

BERT:SETup:DENable
BERT:SETup:IGNore
BERT:SETup:MCOunt
BERT:SETup:MERRor
BERT:SETup:TIMeout
BERT:SETup:TYPE

Cloning the Group

// Create a clone of the original group, that exists independently
var setup2 = driver.Bert.Setup.Clone();

Direct SCPI Members

BertMaskEnum Denable#
// SCPI: BERT:SETup:DENable
BertMaskEnum value = driver.Bert.Setup.Denable;
driver.Bert.Setup.Denable = BertMaskEnum.HIGH;

Activates the Data Enable signal and, if activated, sets its polarity. The signal marks the data that is evaluated by the BERT/BLER measurement. Does not evaluate data supplied by the DUT, that is additional to the PRBS sequence (e.g. sync, preambles, other channels, etc.) .

denable: HIGH | LOW

  • OFF: Ignore any signal at the data enable input; use all data at the BERT/BLER data input for the measurement.

  • HIGH: Use the data enable signal.Only measures the data at the BERT/BLER data input during a high level of the data enable signal. Interrupts the measurement during a low level.

  • LOW: Use the data enable signal.Only measures the data at the BERT/BLER data input during a low level of the data enable signal. Interrupts the measurement during a high level.


BertPattIgnEnum Ignore#
// SCPI: BERT:SETup:IGNore
BertPattIgnEnum value = driver.Bert.Setup.Ignore;
driver.Bert.Setup.Ignore = BertPattIgnEnum.OFF;

Activates ignoring of pure “0” or “1” bit sequences of at least 32 bits, so that faulty frames are excluded from the measurement. Some mobile radio standards indicate faulty frames with pure “0” or “1” bit sequences when errors (e.g. an incorrect checksum) are detected.

ignore: OFF | ZERO | ONE

  • OFF: Pattern ignore is not active.

  • ZERO: Ignore bit sequences of 32 or more consecutive “0” for the BERT measurement.

  • ONE: Ignore bit sequences of 32 or more consecutive “1” for the BERT measurement.


int Mcount#
// SCPI: BERT:SETup:MCOunt
int value = driver.Bert.Setup.Mcount;
driver.Bert.Setup.Mcount = 1;

Enters the number of transmitted data bits/data blocks to be checked before the measurement is terminated. A BERT/BLER measurement does not count data suppressed by method RsSmw.Bert.Setup.Denable | method RsSmw.Bler.Setup.Denable. This termination criteria always terminates the measurement after the specified number of data bits/data blocks. Starting from this point, outputs the fourth value with 1 (= terminate measurement) if the result is queried. For the continuous measurement mode, the measurement restarts once the results have been queried.

mcount: integer


int Merror#
// SCPI: BERT:SETup:MERRor
int value = driver.Bert.Setup.Merror;
driver.Bert.Setup.Merror = 1;

Enters the number of errors to occur before the measurement is terminated. This termination criterion always terminates the measurement after the specified number of errors. Starting from this point, outputs the fourth value with 1 (= terminate measurement) if the measurement result is queried.

merror: integer


double Timeout#
// SCPI: BERT:SETup:TIMeout
double value = driver.Bert.Setup.Timeout;
driver.Bert.Setup.Timeout = 1.0;

Sets the timeout.

timeout: float


BertPrbsEnum Type#
// SCPI: BERT:SETup:TYPE
BertPrbsEnum value = driver.Bert.Setup.Type;
driver.Bert.Setup.Type = BertPrbsEnum.PN11;

Selects the PRBS sequence. The data generated by the PRBS generator is used as a reference for the measurement.

type: PRBS9 | PRBS11 | PRBS15 | PRBS16 | PRBS20 | PRBS21 | PRBS23


Subgroups