from .......Internal.Core import Core
from .......Internal.CommandsGroup import CommandsGroup
from .......Internal import Conversions
from ....... import repcap
# noinspection PyPep8Naming,PyAttributeOutsideInit,SpellCheckingInspection
[docs]class PmOffsetCls:
	"""PmOffset commands group definition. 1 total commands, 0 Subgroups, 1 group commands"""
	def __init__(self, core: Core, parent):
		self._core = core
		self._cmd_group = CommandsGroup("pmOffset", core, parent)
[docs]	def set(self, pm_offset: float, probe=repcap.Probe.Default) -> None:
		"""SCPI: [SENSe]:PROBe<pb>:SETup:PMOFfset \n
		Snippet: driver.applications.k91Wlan.sense.probe.setup.pmOffset.set(pm_offset = 1.0, probe = repcap.Probe.Default) \n
		Sets the P-mode offset. The setting is only available if a modular probe in P-mode is connected to the FSW. The maximum
		voltage difference between the positive and negative input terminals is 16 V. If the probe is disconnected, the P-mode
		offset of the probe is reset to 0.0 V. Note that if the offset for DM-mode or CM-mode is changed, the offsets for the
		P-mode and N-mode are adapted accordingly, and vice versa. For details see 'Using Probes'. \n
			:param pm_offset: The voltage offset between the positive input terminal and ground. Unit: V
			:param probe: optional repeated capability selector. Default value: Nr1 (settable in the interface 'Probe')
		"""
		param = Conversions.decimal_value_to_str(pm_offset)
		probe_cmd_val = self._cmd_group.get_repcap_cmd_value(probe, repcap.Probe)
		self._core.io.write(f'SENSe:PROBe{probe_cmd_val}:SETup:PMOFfset {param}') 
[docs]	def get(self, probe=repcap.Probe.Default) -> float:
		"""SCPI: [SENSe]:PROBe<pb>:SETup:PMOFfset \n
		Snippet: value: float = driver.applications.k91Wlan.sense.probe.setup.pmOffset.get(probe = repcap.Probe.Default) \n
		Sets the P-mode offset. The setting is only available if a modular probe in P-mode is connected to the FSW. The maximum
		voltage difference between the positive and negative input terminals is 16 V. If the probe is disconnected, the P-mode
		offset of the probe is reset to 0.0 V. Note that if the offset for DM-mode or CM-mode is changed, the offsets for the
		P-mode and N-mode are adapted accordingly, and vice versa. For details see 'Using Probes'. \n
			:param probe: optional repeated capability selector. Default value: Nr1 (settable in the interface 'Probe')
			:return: pm_offset: The voltage offset between the positive input terminal and ground. Unit: V"""
		probe_cmd_val = self._cmd_group.get_repcap_cmd_value(probe, repcap.Probe)
		response = self._core.io.query_str(f'SENSe:PROBe{probe_cmd_val}:SETup:PMOFfset?')
		return Conversions.str_to_float(response)