[docs]classSymbolRateCls:"""SymbolRate commands group definition. 1 total commands, 0 Subgroups, 1 group commands"""def__init__(self,core:Core,parent):self._core=coreself._cmd_group=CommandsGroup("symbolRate",core,parent)
[docs]defset(self,sample_rate:float)->None:"""SCPI: SYSTem:COMMunicate:RDEVice:OSCilloscope:SRATe \n Snippet: driver.applications.k6Pulse.system.communicate.rdevice.oscilloscope.symbolRate.set(sample_rate = 1.0) \n Determines whether the 10 GHz mode (default) or 20 GHz mode of the connected oscilloscope is used. The 20 GHZ mode achieves a higher decimation gain, but reduces the record length by half. \n :param sample_rate: 10 GHz | 20 GHz No other sample rate values are allowed. Unit: HZ """param=Conversions.decimal_value_to_str(sample_rate)self._core.io.write(f'SYSTem:COMMunicate:RDEVice:OSCilloscope:SRATe {param}')
[docs]defget(self)->float:"""SCPI: SYSTem:COMMunicate:RDEVice:OSCilloscope:SRATe \n Snippet: value: float = driver.applications.k6Pulse.system.communicate.rdevice.oscilloscope.symbolRate.get() \n Determines whether the 10 GHz mode (default) or 20 GHz mode of the connected oscilloscope is used. The 20 GHZ mode achieves a higher decimation gain, but reduces the record length by half. \n :return: sample_rate: No help available"""response=self._core.io.query_str(f'SYSTem:COMMunicate:RDEVice:OSCilloscope:SRATe?')returnConversions.str_to_float(response)