Source code for RsFsw.Implementations.Applications.K149_Uwb.Trigger.Sequence.Dtime

from ......Internal.Core import Core
from ......Internal.CommandsGroup import CommandsGroup
from ......Internal import Conversions


# noinspection PyPep8Naming,PyAttributeOutsideInit,SpellCheckingInspection
class DtimeCls:
	"""
	| Commands in total: 1
	| Subgroups: 0
	| Direct child commands: 1
	"""

	def __init__(self, core: Core, parent):
		self._core = core
		self._cmd_group = CommandsGroup("dtime", core, parent)

[docs] def set(self, time: float) -> None: """ ``TRIGger[:SEQuence]:DTIMe`` \n Snippet: ``driver.applications.k149Uwb.trigger.sequence.dtime.set(time = 1.0)`` \n Defines the time the input signal must stay below the trigger level before a trigger is detected again. For input from the 'Analog Baseband' interface using the baseband power trigger (BBP) , the default drop out time is set to 100 ns to avoid unintentional trigger events (as no hysteresis can be configured in this case) . :param time: Dropout time of the trigger. Range: 0 s to 10.0 s , Unit: S """ param = Conversions.decimal_value_to_str(time) self._core.io.write(f'TRIGger:SEQuence:DTIMe {param}')
[docs] def get(self) -> float: """ ``TRIGger[:SEQuence]:DTIMe`` \n Snippet: ``value: float = driver.applications.k149Uwb.trigger.sequence.dtime.get()`` \n Defines the time the input signal must stay below the trigger level before a trigger is detected again. For input from the 'Analog Baseband' interface using the baseband power trigger (BBP) , the default drop out time is set to 100 ns to avoid unintentional trigger events (as no hysteresis can be configured in this case) . """ response = self._core.io.query_str(f'TRIGger:SEQuence:DTIMe?') return Conversions.str_to_float(response)