IqPower

SCPI Command :

TRIGger[:SEQuence]:LEVel:IQPower
class IqPowerCls[source]

IqPower commands group definition. 1 total commands, 0 Subgroups, 1 group commands

get() float[source]
# SCPI: TRIGger[:SEQuence]:LEVel:IQPower
value: float = driver.trigger.sequence.level.iqPower.get()

Defines the magnitude the I/Q data must exceed to cause a trigger event. Note that any RF attenuation or preamplification is considered when the trigger level is analyzed. If defined, a reference level offset is also considered.

return

trigger_level: Range: -130 dBm to 30 dBm, Unit: DBM

set(trigger_level: float) None[source]
# SCPI: TRIGger[:SEQuence]:LEVel:IQPower
driver.trigger.sequence.level.iqPower.set(trigger_level = 1.0)

Defines the magnitude the I/Q data must exceed to cause a trigger event. Note that any RF attenuation or preamplification is considered when the trigger level is analyzed. If defined, a reference level offset is also considered.

param trigger_level

Range: -130 dBm to 30 dBm, Unit: DBM