RefLevel#

SCPI Command :

[SENSe]:LIST:RANGe<ri>:RLEVel
Commands in total: 1
Subgroups: 0
Direct child commands: 1
get ( rangePy=RangePy.Default ) → float
set ( ref_level: float, rangePy=RangePy.Default ) → None
# [SENSe]:LIST:RANGe<ri>:RLEVel
# get:
value: float = driver.sense.listPy.range.refLevel.get(rangePy = repcap.RangePy.Default)
# set:
driver.sense.listPy.range.refLevel.set(ref_level = 1.0, rangePy = repcap.RangePy.Default)

No help available

Parameters:
  • ref_level – -130 dBm to 30 dBm (–10 dBm + RF attenuation – RF preamplifier gain)

  • rangePy – optional repeated capability selector. Default value: Ix1 (settable in the interface ‘Range’)