Auto

SCPI Command :

[SENSe]:ESPectrum<sb>:RANGe<ri>:INPut:SANalyzer:ATTenuation:AUTO
class AutoCls[source]

Auto commands group definition. 1 total commands, 0 Subgroups, 1 group commands

set(state: bool, subBlock=SubBlock.Default, rangePy=RangePy.Default) None[source]
# SCPI: [SENSe]:ESPectrum<sb>:RANGe<ri>:INPut:SANalyzer:ATTenuation:AUTO
driver.sense.espectrum.range.inputPy.sanalyzer.attenuation.auto.set(state = False, subBlock = repcap.SubBlock.Default, rangePy = repcap.RangePy.Default)

This command turns automatic selection of the input attenuation for a SEM range on and off. If an external frontend is active (see [SENSe:]EFRontend[:STATe]) , you can configure attenuation settings for the external frontend and the analyzer separately. This setting defines the attenuation for the analyzer. In case of high speed measurements, the input attenuation has to be identical for all ranges.

param state

ON | OFF | 0 | 1

param subBlock

optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Espectrum’)

param rangePy

optional repeated capability selector. Default value: Ix1 (settable in the interface ‘Range’)