Mode#

SCPI Command :

CALCulate<n>:DELTamarker<m>:FUNCtion:BPOWer:MODE
Commands in total: 1
Subgroups: 0
Direct child commands: 1
get ( window=Window.Default, deltaMarker=DeltaMarker.Default ) → MarkerMode
set ( mode: MarkerMode, window=Window.Default, deltaMarker=DeltaMarker.Default ) → None
# CALCulate<n>:DELTamarker<m>:FUNCtion:BPOWer:MODE
# get:
value: enums.MarkerMode = driver.calculate.deltaMarker.function.bpower.mode.get(window = repcap.Window.Default, deltaMarker = repcap.DeltaMarker.Default)
# set:
driver.calculate.deltaMarker.function.bpower.mode.set(mode = enums.MarkerMode.DENSity, window = repcap.Window.Default, deltaMarker = repcap.DeltaMarker.Default)

Selects the way the results for a band power delta marker are displayed.

Parameters:
  • mode –

    • POWer: Result is displayed as an absolute power. The power unit depends on the CALCulaten:UNIT:POWer setting.

    • DENSity: Result is displayed as a density in dBm/Hz.

    • RPOWer: This setting is only available for a delta band power marker.The result is the difference between the absolute power in the band around the delta marker and the absolute power for the reference marker. The powers are subtracted logarithmically, so the result is a dB value.[Relative band power (Delta2) in dB] = [absolute band power (Delta2) in dBm] - [absolute (band) power of reference marker in dBm]For details see ‘Relative band power markers’.

  • window – optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Calculate’)

  • deltaMarker – optional repeated capability selector. Default value: Nr1 (settable in the interface ‘DeltaMarker’)